Product Introduction
Multi-clinical usage:
Neuro Intervention, Cardiac Intervention, Oncology Intervention and Hybird OR
Use Environment:
-
Power requirements: three-phase AC
-
voltage: 380V±10% frequency: 50/60Hz±1Hz
-
Ambient temperature:10℃-35℃
- Relative humidity: 30%-75%
Angiography Machine Index:
High-speed Rotating Anode Tube:
-
Anode maximum dissipation: 4.5kW
-
Dual focus: small 0.6mm, large 1.0mm
-
Tube cooling method: Air, Water, Oil
-
Anode rotate speed 9000r/min
-
Small focus power: 43KW
-
Big focus power: 80KW
-
Anode target material: Molybdenum rhenium
-
Target angle:11°
The X Ray Generator:
-
Power: 100kW
-
Maximum output voltage:125KV
-
Exposure time: 1ms-6300ms
-
Exposure current time product: 0.1mAs-1000mAs
-
Automatic exposure control without test exposure
-
Variable frequency high voltage generator, inverter frequency: 160KHz-400KHz
The Detector:
-
The collimator and flat panel detector synchronic servo ensure the clinic imaging is always upright
-
Flat panel detector rotation: ±90°
-
Digital imaging acquisition matrix: 1536×1536
-
Maximum acquisition rate:30fps
-
Nominal area: 30cm × 30cm
-
A/D range: 16bit
The 7-axis Gantry:
-
Easy to operate and learn
-
One-key emergency stop
-
R-axis rotation range: 300°
-
P-axis rotation range : 316°
-
O-axis rotation range: 256°
Image Acquisition:
-
Dynamic image optimization and noise reduction
-
Adaptive edge enhancement
-
Window width/window level adjustment, sharpness adjustment, Gamma adjustment
-
Last image hold
Automatic collimator:
- Automatic switching filter: Air, 0.1mmCu, 1mmAl+0.2mmCu, 2mmAl
-
Electric lung field iris
Image Display Function:
-
Acquisition time
-
Date display
-
Image freezing
-
Grayscale inversion
-
Image annotation
-
Left/right identification
-
Text annotation
-
Anatomical background
Image Post-processing System (option):
-
XA-M intervention: The workstation can simultaneously receive, display, and process international standard images of CT, MR, US, and other types of equipment, as well as calibration, measurement, and image processing of related image
-
Left ventricular ejection analysis, and vascular stenosis analysis, provide parameters of cardiac output, ejection fraction, wall motion analysis, and vascular stenosis analysis
-
Post-processing workstation hardware configuration
Angiography Machine
Clinical Image
Hot Tags: intelligent fpd floor-mounted cath lab, suppliers, factory, pricelist, Italy, Europe, CE, for sale, intervention, pci operation, cardiac vascular angiography machine, digital x ray system, interventional radiology gynecology, Medical Dsa System
You Might Also Like